THE STUDY OF METHODS FOR MEASURING TISSUE OXYGENATION BY USING NIRS
DOI:
Author:
Affiliation:

Clc Number:

R329.25 TN219

Fund Project:

  • Article
  • |
  • Figures
  • |
  • Metrics
  • |
  • Reference
  • |
  • Related
  • |
  • Cited by
  • |
  • Materials
  • |
  • Comments
    Abstract:

    Based on the model of multilayered biological tissue, a method to measure tissue oxygenation saturation by using noninvasive near infrared spectroscopy was present. For the subjects with different constructions and scales of their overlaying tissue, the correction method which make results reliable and comparable in semi quantitative measuring change in tissue blood oxygenation was proposed. A new compact monitoring instrument which includes a probe with a dual wavelength LED and two integrated optical detectors at different distance from light source was designed and developed. During the ischaemia of forearm induced by 240mmHg pressure with inflating cuff, the result of concentration changes in oxyhemoglobin, deoxyhemoglobin and the value of tSO 2 is consistent with the foreign reports by TRS methods. It shows that the monitoring system is very stable and feasible.

    Reference
    Related
    Cited by
Get Citation

HUANG Lan TIAN Feng Hua DING Hai Shu WANG Guang Zhi TENG Yi Chao. THE STUDY OF METHODS FOR MEASURING TISSUE OXYGENATION BY USING NIRS[J]. Journal of Infrared and Millimeter Waves,2003,22(5):379~383

Copy
Share
Article Metrics
  • Abstract:
  • PDF:
  • HTML:
  • Cited by:
History
  • Received:
  • Revised:September 19,2002
  • Adopted:
  • Online:
  • Published: