基于激光检测提高复合面源排放测量精度研究
作者:
作者单位:

1.中国科学院环境光学与技术重点实验室,安徽光学精密机械研究所,安徽 合肥 230031;2.安徽农业大学,安徽 合肥 230061;3.合肥工业大学,安徽 合肥 230009

作者简介:

通讯作者:

中图分类号:

O43

基金项目:


Improving measurement accuracy of composite non-point sources emissions based on laser detection
Author:
Affiliation:

1.Key Laboratory of Environmental Optics& Technology, Anhui Institute of Optics and Fine Mechanics, the Chinese Academy of Sciences, Hefei, 230031, China;2.Anhui Agricultural University, Hefei, 230061, China;3.Hefei University of Technology, Hefei 230009, China

Fund Project:

Supported by the National Natural Science Foundation of China (41805124); Chinese Academy of Sciences Scientific Research Instrument and Equipment Development Project(YJKYYQ20190051) Key Research and Development Plan of Anhui Province (201904a07020093); Anhui Natural Science Foundation Project (1808085MD107).National Engineering Laboratory of Advanced Technology and Equipment for Atmospheric Pollution Monitoring(2005DP173065-2021-04)

  • 摘要
  • |
  • 图/表
  • |
  • 访问统计
  • |
  • 参考文献
  • |
  • 相似文献
  • |
  • 引证文献
  • |
  • 资源附件
  • |
  • 文章评论
    摘要:

    在面源气体排放的测量过程中,不可避免地受到相邻田地气体排放干扰,又受到环境扰动的影响,影响气体检测的准确性。为了提高测量精度,研究了基于激光光谱检测的复合面源甲烷非均匀排放测量方法。通过多次自标定测量,减少了外部环境的干扰。设计了非均匀面源排放物检测系统,并提出了消除气体波动检测方法。首先进行了准确性验证试验,待测源a的标准差为0.17,待测源b的标准差为0.18;随后与光声光谱抽取式进行了对比试验,相关系数达到0.91;最后开展了外场实际的测量应用,对不同施肥方式造成的两个不均匀排放待测源进行监测,实现对多源不均匀排放的准确测量。它具有农业气体排放,环境气体检测等实用工程价值。

    Abstract:

    In the measurement of source emissions, it is inevitably interfered by the gas emissions from the adjacent fields and turbulence, which affects the accuracy of gas detection. In order to improve the measurement accuracy, the measurement method of composite non-point sources non-uniformity methane emissions based on laser spectrum detection has been studied. Moreover, the interference in external environment is reduced with multiple self-calibration measurements. A detection system for composite non-point source emissions has been established, and the detection method of eliminating gas fluctuation has been proposed. First, the accuracy verification test has been carried out. The standard deviation of source a is 0.17, and that of source b is 0.18. Subsequently, a comparative test has been carried out with the extraction method of the photo acoustic spectrum, and the correlation coefficient reached 0.91. Finally, the actual field measurement has been carried out to monitor the two non-uniform emission sources caused by different fertilization methods to achieve accurate measurement. It has practical engineering value such as agricultural gas emission and environmental gas detection.

    参考文献
    相似文献
    引证文献
引用本文

唐七星,张玉钧,何莹,王立明,李梦琪,尤坤,李潇毅,陈东,刘文清.基于激光检测提高复合面源排放测量精度研究[J].红外与毫米波学报,2021,40(3):363~368]. TANG Qi-Xing, ZHANG Yu-Jun, HE Ying, WANG Li-Ming, LI Meng-Qi, YOU Kun, LI Xiao-Yi, CHEN Dong, LIU Wen-Qing. Improving measurement accuracy of composite non-point sources emissions based on laser detection[J]. J. Infrared Millim. Waves,2021,40(3):363~368.]

复制
分享
文章指标
  • 点击次数:
  • 下载次数:
  • HTML阅读次数:
  • 引用次数:
历史
  • 收稿日期:2020-07-18
  • 最后修改日期:2021-04-29
  • 录用日期:2020-08-24
  • 在线发布日期: 2021-04-27
  • 出版日期: 2021-06-25