制冷型红外探测器杜瓦辐射漏热的精确计算
CSTR:
作者:
作者单位:

昆明物理研究所,云南昆明 650223

作者简介:

通讯作者:

中图分类号:

基金项目:


Precise calculation of radiation heat of cryogenic infrared detector Dewar
Author:
Affiliation:

Kunming Institute of Phyics, Kunming 650223,China

Fund Project:

  • 摘要
  • |
  • 图/表
  • |
  • 访问统计
  • |
  • 参考文献
  • |
  • 相似文献
  • |
  • 引证文献
  • |
  • 资源附件
  • |
  • 文章评论
    摘要:

    制冷型红外探测器杜瓦的热负载是表征杜瓦绝热能力的综合指标,辐射漏热是热负载的一部分,计算辐射漏热时,传统的方法一般将杜瓦简化为同轴圆筒模型,简化模型与实际模型相差较大且无法计算发射率、透射率和反射率随波长变化的表面之间的辐射换热。为提高杜瓦辐射漏热的计算精度,基于蒙特卡洛原理,采用3D Studio Max建模,提取模型信息开发程序,得到了一套基于辐射传递因子的杜瓦辐射漏热通用计算程序。为了初步检验计算程序的准确性,按灰体假设计算了两型实验杜瓦的冷端辐射漏热并与实验测量值进行比对,实验杜瓦1(典型1K×1K)冷端辐射漏热的理论计算值和实验测量值分别为155mW和136mW,两者的误差为19mW;实验杜瓦2(典型640×512)冷端辐射漏热的理论计算值和实验测量值分别为87mW和79mW,两者的误差为8mW。初步检验计算程序的准确性后,考虑材料表面发射率、透射率和反射率随波长的变化及温度对辐射波长的影响,计算和实验测量了工程用典型1K1K长波杜瓦在窗片面对发射率为0.9的灰体时的冷端辐射漏热,理论计算值为127mW,实验测量值为110mW,两者的误差为17mW。

    Abstract:

    The thermal load of the cryogenic infrared detector Dewar is a comprehensive indicator characterizing the adiabatic capacity of the Dewar. Radiative heat is a part of the thermal load. When calculating the radiative heat ,the traditional approach typically simplifies the Dewar to a coaxial cylindrical model. This simplified model differs significantly from the actual one and the traditional approach is incapable of computing the radiative heat transfer between surfaces where emissivity, transmittance, and reflectance vary with wavelength. To enhance the calculation accuracy of the Dewar"s radiative heat, based on the Monte Carlo principle, a 3D Studio Max modeling was employed, model information was extracted, and a program was developed, resulting in a set of general calculation programs for the Dewar"s radiative heat based on the radiation transfer factor. To preliminarily verify the accuracy of the calculation program, the cold side radiative heat of two types of experimental dewars was calculated according to the gray body assumption and compared with the measured values. The theoretical calculated value and measured value of the cold side radiation heat of experimental dewar 1 were 155 mW and 136 mW, respectively, with an error of 19 mW; the theoretical calculated value and measured value of cold side radiation heat of experimental dewar 2 were 87 mW and 79 mW, respectively, with an error of 8 mW. After initially testing the accuracy of the calculation program, the cold side radiative heat of an engineering typical 1K×1K long-wave dewar when the emissivity of the window facing it was 0.9 was calculated and measured experimentally. The theoretical calculated value was 127 mW, and the measured value was 110 mW, with an error of 17 mW.

    参考文献
    相似文献
    引证文献
引用本文
分享
文章指标
  • 点击次数:
  • 下载次数:
  • HTML阅读次数:
  • 引用次数:
历史
  • 收稿日期:2024-11-05
  • 最后修改日期:2024-12-13
  • 录用日期:2024-12-18
  • 在线发布日期: 2025-02-18
  • 出版日期:
文章二维码