Two-point calibration technique of microwave radiometer receiver was presented and its errors were analyzed. 23.8GHz digital auto gain compensative microwave radiometer was calibrated and the calibration error range was calculated then. The accuracy of the calibration equation was verified by experiment. The results show that two-point calibration has favorable veracity for microwave radiometer receiver calibration.
参考文献
相似文献
引证文献
引用本文
栾卉,赵凯.微波辐射计接收机两点定标法误差分析及准确性验证[J].红外与毫米波学报,2007,26(4):289~292]. LUAN Hui, ZHAO Kai. ERROR ANALYSIS AND ACCURACY VALIDATION OF TWO-POINT CALIBRATION FOR MICROWAVE RADIOMETER RECEIVER[J]. J. Infrared Millim. Waves,2007,26(4):289~292.]