FY-2D红外探测器窗区通道非线性特性及其定标订正方案
DOI:
作者:
作者单位:

国家气象局国家卫星气象中心,国家气象局国家卫星气象中心,国家气象局国家卫星气象中心,国家气象局国家卫星气象中心

作者简介:

通讯作者:

中图分类号:

基金项目:

国家杰出青年科学基金


Nonlinearity of FY-2D infrared detector in thermal window channels and its correction method
Author:
Affiliation:

National Satellite Meteorological Center,National Satellite Meteorological Center,National Satellite Meteorological Center,National Satellite Meteorological Center

Fund Project:

  • 摘要
  • |
  • 图/表
  • |
  • 访问统计
  • |
  • 参考文献
  • |
  • 相似文献
  • |
  • 引证文献
  • |
  • 资源附件
  • |
  • 文章评论
    摘要:

    利用实验室定标与基于高光谱探测器IASI的交叉匹配样本,分析了FY-2D红外探测器在窗区通道的非线性特征,结果显示,VISSR在信号两端都存在非线性特征,但对低信号端的影响最为显著.与实验室测量结果比较,线性定标在低温端会带来大于10 K的亮温偏差.为了同时保证定标结果的质量及其稳定性,采用分段定标方案,高温端为线性定标而中低温端为非线性.从一致性以及台风个例分析等方面对定标方案进行了验证,分析结果表明,分段定标方案稳定,对低温端的定标精度改进显著,与线性结果比较台风监测亮温精度提升大于7K.

    Abstract:

    FY-2 Visible and Infrared Spin Scan Radiometer had obviously nonlinear characteristics at low signal. It brought significant biases when using linear calibration in low temperature targets monitoring especially for typhoon targets, which seriously influences FY-2 quantitative application. Nonlinearity of FY-2D infrared detector in window channels was investigated using laboratory calibration measurements and collocation samples based on high-spectral sensor IASI. It was found that nonlinear performance existed both at high and low signal, but impacts at low signal were the most significant. Comparing with laboratory measurements, linear calibration could cause more than 10 K deviations at low temperature. In order to guarantee the quality of calibration results and its stability, nonlinear calibration scheme was applied only in low temperature area. Nonlinear calibration scheme was investigated from its consistency, and also typhoon application. The results indicated nonlinear calibration was stable, and significantly improved calibration accuracy at low temperature, especially in typhoon monitoring which was advanced at least by 7K.

    参考文献
    相似文献
    引证文献
引用本文

徐娜,陈林,胡秀清,陆风. FY-2D红外探测器窗区通道非线性特性及其定标订正方案[J].红外与毫米波学报,2013,32(4):337~343]. XU Na, CHEN Lin, Hu Xiu-Qing, LU Feng. Nonlinearity of FY-2D infrared detector in thermal window channels and its correction method[J]. J. Infrared Millim. Waves,2013,32(4):337~343.]

复制
分享
文章指标
  • 点击次数:
  • 下载次数:
  • HTML阅读次数:
  • 引用次数:
历史
  • 收稿日期:2012-03-21
  • 最后修改日期:2012-11-10
  • 录用日期:2012-05-16
  • 在线发布日期: 2013-08-29
  • 出版日期:
文章二维码