High-efficiency high-accuracy data acquisition method and realization for IRFPA test system
DOI:
CSTR:
Author:
Affiliation:

Clc Number:

Fund Project:

  • Article
  • |
  • Figures
  • |
  • Metrics
  • |
  • Reference
  • |
  • Related
  • |
  • Cited by
  • |
  • Materials
  • |
  • Comments
    Abstract:

    The data requirements of all parameters on the performance test of infrared focal plane arrays (IRFPA) was analyzed, and then presented an efficient data acquisition method different from the traditional one. As a result of ADI latest high-performance 16-bit ADC and this new method, a high-efficiency high-accuracy data acquisition system was introduced. By the embedded RMS Hardware Algorithm and the particular combination of external SRAMs, this scheme is not only compatible with conventional test system, but also improves the efficiency of the test system greatly at the same time. Through experiments on HgCdTe 640?512(15 μm) IRFPA, the noise floor of the acquisition system is down to 75 μVrms and the calculation of various parameters can be completed in about 2 seconds. The data acquisition system is proved to be able to meet the demanding requirements for full IRFPA test and will have a great application prospect relatively to the traditional one.

    Reference
    Related
    Cited by
Get Citation

HUANG Da-Jun, WANG Yong-Gang, ZHANG Li-Jun, ZHU Wen-Song, LU Xiao-Ming. High-efficiency high-accuracy data acquisition method and realization for IRFPA test system[J]. Journal of Infrared and Millimeter Waves,2011,30(4):305~309

Copy
Share
Article Metrics
  • Abstract:
  • PDF:
  • HTML:
  • Cited by:
History
  • Received:October 12,2010
  • Revised:April 02,2011
  • Adopted:January 30,2011
  • Online: August 25,2011
  • Published:
Article QR Code