基于集成黑体法的1 000~1 500℃石墨材料高温红外光谱发射率测量
作者:
作者单位:

中国计量科学研究院 热工所,北京 100029

作者简介:

通讯作者:

中图分类号:

TB942

基金项目:

国家自然科学基金(11772318)


The infrared spectral emissivity measurement of a graphite material in a high temperature range of 1000~1500℃ using integrated blackbody principle
Author:
Affiliation:

Heat Division, National Institute of Metrology, Beijing 100029, China

Fund Project:

Supported by National Natural Science Foundation of China (11772318)

  • 摘要
  • |
  • 图/表
  • |
  • 访问统计
  • |
  • 参考文献
  • |
  • 相似文献
  • |
  • 引证文献
  • |
  • 资源附件
  • |
  • 文章评论
    摘要:

    建立了基于集成黑体法的材料高温红外光谱发射率测量理论模型,重点讨论了不等温集成黑体空腔有效发射率、观测因子、样品材料外推温降因素的影响。基于集成黑体法搭建了以傅里叶红外光谱仪为红外辐射探测系统的高温红外光谱发射率测量装置。基于蒙特卡罗光线追迹法数值开展了集成黑体有效发射率数值模拟,并进行了实验验证。讨论了辐射源尺寸效应、光谱响应度线性度等非理想因素对集成黑体法高温光谱发射率测量的影响。开展了石墨材料在1 000℃、1 300℃、1 500℃的红外光谱发射率实验研究,并与已公开发表的测量结果进行了比对。本文结果与文献数据优于5%的良好一致性,验证了集成黑体方法在高温发射率测量中的可行性。

    Abstract:

    A theoretical model of the infrared spectral emissivity measurement of materials using the integrated blackbody principle was established. The effects of the effective emissivity of the non-isothermal integrated blackbody cavity, the observation coefficient and the temperature drop during the sample material push-out were investigated. The device of the infrared spectral emissivity measurement using the integrated blackbody principle was set up which used a Fourier-Transform infrared spectrometer as the infrared radiation detection instrument. The effective emissivity of the integrated blackbody cavity was simulated using Monte-Carlo ray tracing method and the associated validation experiments were carried out. The effects of non-ideal factors, including the size-of-source effect and linearity of the spectral responsivity, on the infrared spectral emissivity measurement using the integrated blackbody principle were investigated. The experimental measurements of the infrared spectral emissivity of a graphite material were carried out at 1000℃, 1300℃ and 1500℃, respectively. The results in this article are in good agreement with the literature data better than 5%, which verifies the feasibility of the emissivity measurement method using the integrated blackbody principle at high temperatures.

    参考文献
    相似文献
    引证文献
引用本文

宋旭尧,董伟,潘奕捷,原遵东,卢小丰.基于集成黑体法的1 000~1 500℃石墨材料高温红外光谱发射率测量[J].红外与毫米波学报,2021,40(2):204~213]. SONG Xu-Yao, DONG Wei, PAN Yi-Jie, YUAN Zun-Dong, LU Xiao-Feng. The infrared spectral emissivity measurement of a graphite material in a high temperature range of 1000~1500℃ using integrated blackbody principle[J]. J. Infrared Millim. Waves,2021,40(2):204~213.]

复制
分享
文章指标
  • 点击次数:
  • 下载次数:
  • HTML阅读次数:
  • 引用次数:
历史
  • 收稿日期:2020-06-23
  • 最后修改日期:2021-04-05
  • 录用日期:2020-09-04
  • 在线发布日期: 2021-03-30
  • 出版日期: