一种新型单片集成光谱传感器的波长定标方法
作者:
作者单位:

山东大学 光学高等研究中心,山东大学 光学高等研究中心,山东大学 光学高等研究中心,山东大学 光学高等研究中心,山东大学 光学高等研究中心,中国科学院上海技术物理研究所传感技术国家重点实验室,山东大学 光学高等研究中心

作者简介:

通讯作者:

中图分类号:

基金项目:

国家自然科学基金面上项目(61376052);传感技术联合国家重点实验室课题(SKT1507)


Wavelength calibration of a new monolithically integrated spectral sensor
Author:
Affiliation:

Advanced Research Center for Optics,Shandong University,Advanced Research Center for Optics,Shandong University,Advanced Research Center for Optics,Shandong University,Advanced Research Center for Optics,Shandong University,Advanced Research Center for Optics,Shandong University,State Key Laboratories of Transducer Technology, Shanghai Institute of Microsystem and Information Technology, Chinese Academy of Sciences,Advanced Research Center for Optics,Shandong University

Fund Project:

  • 摘要
  • |
  • 图/表
  • |
  • 访问统计
  • |
  • 参考文献
  • |
  • 相似文献
  • |
  • 引证文献
  • |
  • 资源附件
  • |
  • 文章评论
    摘要:

    新型短波红外单片集成光谱传感器(MCSS)是光谱传感物联网的核心感知部件.提出了一种MCSS的波长定标方法, 根据相对光谱响应度和波长分配原则把相邻像素组成一个定标通道.分别采用MCSS和两款商用光谱仪测量标准物质的光谱.实验结果表明, 平均定标偏差为1.8 nm, 与一款具有相似FWHM和取样间隔的商用光谱仪接近, 能够满足实际应用的需要

    Abstract:

    A new near-infrared monolithically integrated multichannel spectral sensor (MCSS) serves as a key component of Spectral-sensing Internet of Things (SSIOT). A wavelength calibration method for the MCSS was proposed. The method is based on the relative spectral response and includes specific principles of wavelength allocation. Several pixels comprised a channel with the calibrated wavelength. Experimental tests were conducted in the calibrated MCSS and other two commercial spectrometers. The compared spectrum results demonstrated calibration accuracy of 1.8 nm on average, with performance as good as a commercial spectrometer equipped with similar FWHM and sampling intervals, which can meet the application demand.

    参考文献
    相似文献
    引证文献
引用本文

张亚光,庄新港,王绪泉,刘俊良,方家熊,邵秀梅,王丽丽.一种新型单片集成光谱传感器的波长定标方法[J].红外与毫米波学报,2017,36(1):15~19]. ZHANG Ya-Guang, ZHUANG Xin-Gang, WANG Xu-Quan, LIU Jun-Liang, FANG Jia-Xiong, SHAO Xiu-Mei, WANG Li-Li. Wavelength calibration of a new monolithically integrated spectral sensor[J]. J. Infrared Millim. Waves,2017,36(1):15~19.]

复制
分享
文章指标
  • 点击次数:
  • 下载次数:
  • HTML阅读次数:
  • 引用次数:
历史
  • 收稿日期:2016-04-12
  • 最后修改日期:2016-04-14
  • 录用日期:2016-04-19
  • 在线发布日期: 2017-03-28
  • 出版日期: